25
c.
Press
twice to skip calibration adjustment and accept
the new tempco (3 times if in TDS mode). You are now ready to
measure samples with your new temperature compensation factor.
2.
Disabling Temperature Compensation
a.
Select User mode (ref. Procedure to Select a Solution, pg. 23).
b.
With “User” selected, press
. If the display does not
show .00%/°C, hold
long enough to bring the tempco to
.00%/°C (see Figure 3).
c.
Press
twice (3 times if in TDS mode). Temperature
compensation is now disabled (=0) for measurements in User
mode.
3.
User Programmable Conductivity to TDS Ratio
This feature allows you to select a custom conductivity to TDS conversion
ratio within the range of 0.20-7.99 for User mode measurements.
To determine the conversion ratio for a custom solution of known TDS
ppm value, measure the solution conductivity at 25ºC with the Ultrameter
III
and divide the ppm value by the µS value. For example, a solution of
known 75 ppm TDS and measured 100 µS conductivity at 25ºC would
have a conversion ratio of 75/100 or 0.75. Enter the new conversion
ratio as follows: