16
F. Measuring Hardness
1.
Hardness Unit Selection
The 9P offers the ability to set the hardness unit preference to either
“PPM” CaCO
3
or grains of hardness. To change hardness units, press
HARD, then press and hold CAL down until “HArd” and “SEL” are
displayed. Press UP or DOWN to toggle between “PPM” and grains (no
units are displayed for grains). Press CAL to accept. This also sets the
hardness unit preference for the LSI Calculator simultaneously.
2.
Select EDTA Reagent
0-200 ppm- Reagent H4, EDTA-B-LC (optimal for 0-100 ppm)
0-1710 ppm- Reagent H5, EDTA-B-HC (optimal for 100-1710 ppm)
3.
Hardness Titration Procedure
Prepare the materials required for titration: Reagent: HB, Reagent: H4,
or Reagent: H5, cell extender, T-plunger, 100µL pipette with a clean tip
installed, and one extra clean tip.
NOTE: If you do not complete each titration point within the 3-minute
timeout period, the unit exits the Hardness function. Start the titration
process over.
1.
Ensure pH protective cap is secure to avoid contamination.
2.
Install the cell extender by pushing the base of the cell extender
into the conductivity cell until it is fully seated.
3.
Rinse the conductivity cell cup and cell extender 3 times with
solution to be measured.
4.
Refill cell cup and cell extender with sample solution.
5.
Insert T-plunger into open cell extender until the “T” rests flush
on the lip of the cell extender. This will cause fluid beyond what
is required for titration to overflow the cell extender.
6.
Rotate the T-plunger from side to side, allowing the arms of the
T-plunger to slide along the rim of the cell extender.