28
2.
Calibration Limits
There are calibration limits for all functions for which calibration is
required. The 9P will not allow you to calibrate beyond its calibration
limits. Attempts to calibrate too far up or down from the ideal “FAC” value
stored in the unit by the factory will cause the displayed value to be
replaced with “FAC”. If you accept it (press the CAL key), you will have
the original default factory calibration for this measurement. The need to
calibrate so far out that “FAC” appears indicates a procedural problem,
incorrect standard solution, a very dirty cell cup or an aging pH/ORP
sensor (ref. Troubleshooting Chart, pg. 52).
C. Calibration Procedures
1.
Conductivity or TDS Calibration
a.
Rinse conductivity cell 3 times with proper standard (KCl, NaCl,
or 442
TM
) (ref. Cond/TDS Standard Solutions, pg. 54). For user
calibration see User Calibration Conductivity/TDS below.
b.
Refill conductivity cell with same standard. KCl-7000 shown.
c.
Press
or
, then press
; “CAL” icon will
appear on the display (see Figure 5).
d.
Press
or
to step the displayed value toward the
standard’s value (7032 > 7000) or hold a key down to scroll
rapidly through the reading.
e.
Press
once to confirm new value and end the calibration
sequence for this particular solution type. If another solution
type is also to be measured, change solution type now and
repeat this procedure.
2.
User Calibration Conductivity/TDS
Instrument must be in User mode, see Solution Selection, pg. 23.
a.
Rinse conductivity cell three times with YOUR standard.
b.
Refill conductivity cell with same standard.