A KCl correction of 10 (degrees below 25) x 1.9% indicates the solution
is reading 19% low. Dividing by (.81) yields 1111 microsiemens for a
compensated reading. The difference is 29 out of 1082 = 2.7%.
D. A Chart of Comparative Error
In the range of 1000 µS, the error using KCl on a solution that should
be compensated as NaCl or as 442
TM
, is illustrated in the graph below.
Users wanting to measure natural water based solutions to 1% would
have to alter the internal compensation to the more suitable preloaded
“442
TM
” values or stay close to 25°C. Users who have standardized
to KCl- based compensation may want to stick with it, regardless of
increasing error as you get further from 25°C. The Ultrameter
III
will
provide the repeatability and convertibility of data necessary for relative
values for process control.
E. Other Solutions
A salt solution like seawater or liquid fertilizer acts like NaCl. An internal
correction for NaCl can be selected for greatest accuracy with such
solutions. Many solutions are not at all similar to KCl, NaCl or 442
TM
. A
sugar solution, or a silicate, or a calcium salt at a high or low temperature
may require a “User” value unique for the application to provide readings
close to the true compensated conductivity.