![multicomp pro MP700852 User Manual Download Page 42](http://html1.mh-extra.com/html/multicomp-pro/mp700852/mp700852_user-manual_1854247042.webp)
Page <42>
V1.0
07/01/22
Newark.com/multicomp-pro
Farnell.com/multicomp-pro
sg.element14.com/b/multicomp-pro
Sweep
Sweep Time
Span≥10 Hz
10 ms to 3000 s
Zero Span
20 us to 3000 s
Sweep time uncertainty
span ≥ 100 Hz: 5% (nominal)
zero span (sweep time setting value > 1 ms): 5% (nominal)
Sweep Mode
Continuous, Single
Trigger
Trigger source
free run, video, external
External trigger level
5 V TTL level
Tracking Generator (Option)
Frequency Range
XSA805 (TG)
100 kHz to 500.009 MHz
XSA810 (TG)
100 kHz to 1.000009 GHz
Output power level range
-40 dBm to 0 dBm
Output power level resolution 1 dB
Output flatness
relative to 50 MHz
±3 dB
Maximum safe reverse level
Harmonic spuri-
ous
-30 dBc (Tracking generator output power = -10 dBm)
Non-harmonic
spurious
-40 dBc(Tracking generator output power = -10 dBm)
Tracking generator to input
terminal isolation
-60 dB (Tracking generator output power = 0 dBm)
Input/Output
RF Input
Impedance
50 Ω, typical
Connector
N Type Female
Tracking generator output
Impedance
50 Ω, typical
Connector
N Type Female
Internal reference
frequency
10 MHz
output level
+3 dBm to +10 dBm, +8 dBm (typical)
Impedance
50 Ω, typical
Connector
BNC female
External reference
frequency
10 MHz ± 5 ppm
output level
0 dBm to + 10 dBm
Impedance
50 Ω, typical
Connector