Two ways to enable a certain point as the reference:
a)
Open a “Normal” marker and locate it onto a point and then switch the marker type into “Delta”, creating a new
reference, then you can modify the location of the delta point to achieve the delta measurement.
b)
Open a Delta Marker and place it onto a point, then reselect the Delta menu to locate the marker you opened onto this
points, then you can modify the location of the delta point to achieve the delta measurement.
5.14.5 [Off]
The marker information displayed on the screen and functions based on the marker will be turned off and won’t show up
again.
5.14.6 [All Off]
Turns off all the opened markers and the related functions. The marker won’t show again.
5.14.7 [Marker Table On Off]
Turns on or off the display of all marker table.
5.15 [Marker→]
A soft menu associated with the marker function is popped out for setting the other system parameters (such as Center
frequency, Reference level) by current marker readings. These menus relate to the frequency of the spectrum analyzer,
whether the sweep width and marker are in normal or delta marker mode.
5.15.1 [Mkr->CF]
Sets the center frequency of the analyzer based on the frequency of the current marker. This feature quickly moves the signal
to the center of the screen.
•
If Normal is selected, the center frequency will be set to the frequency of the current marker.
•
If Delta Marker is selected, the center frequency will be set to the frequency at which the Delta Marker is located.
• The function is invalid in Zero span mode.
5.15.2 [Mkr->CF Step]
Sets the center frequency step of the analyzer based on the frequency of the current marker.
•
If Normal is selected, the center frequency step will be set to the frequency of current marker.
•
If Delta Marker is selected, the center frequency step will be set to the frequency at which the Delta Marker is located.
• The function is invalid in Zero span mode.
5.15.3 [Mkr->Start]
Sets the start frequency of the analyzer based on the frequency of the current marker.
•
If Normal is selected, the start frequency will be set to the frequency of the current marker.
•
If Delta Marker is selected, the start frequency will be set to the frequency at which the Delta Marker is located.
• The function is invalid in Zero span mode.
5.15.4 [Mkr->Stop]
Sets the stop frequency of the analyzer based on the frequency of the current marker.
•
If Normal is selected, the stop frequency will be set to the frequency of the current marker.
•
If Delta Marker is selected, the stop frequency will be set to the frequency at which the Delta Marker is located.
• The function is invalid in Zero span mode.
5.15.5 [Mkr->Ref Level]
Sets the reference level of the analyzer based on the amplitude of the current marker.
•
If Normal is selected, the reference level will be set to the amplitude of the current marker.
•
If Delta Marker is selected, the reference level will be set to the amplitude at which the Delta Marker is located.
5.15.6 [MkrΔ->Span]
Changes the span of the analyzer to the frequency difference between the two markers.