![multicomp pro MP700852 User Manual Download Page 15](http://html1.mh-extra.com/html/multicomp-pro/mp700852/mp700852_user-manual_1854247015.webp)
Page <15>
V1.0
07/01/22
Newark.com/multicomp-pro
Farnell.com/multicomp-pro
sg.element14.com/b/multicomp-pro
To clearly observe the signal, reduce the frequency span to 1 MHz and set the center frequency to 100MHz.
2) Setting Center Frequency
Press “FREQ”, select [Center frequency] on corresponding pop up menu. Input “100” and select the unit as MHz on the
numeric keypad. The keys can be used to set the exact value but the knob and directional keys can also be used to set the
center frequency.
3) Setting Frequency Span
Press [SPAN], input “
1
” and select
MHz
as its unit using the numeric keypad; or press [↓] to decrease to 1MHz.
Press [BW], set [resolution bandwidth] to manual, and input
“30”
and select
kHz
as its unit using the numeric keypad; or
press [↓] to decrease to 30kHz.
Press [Detector], set the detection type to positive peak.
Figure 4-9 shows the signal at a higher resolution.
Please note that resolution bandwidth, video bandwidth and frequency span are self-adapted. They adjust to certain values
according to frequency span. Sweep time can be self-adapted too.