6. Specification
This chapter lists the technical specifications and general technical specifications of the spectrum analyzer. Unless otherwise
stated, the technical specifications apply to the following conditions:
• The instrument has been preheated for 30 minutes before use.
• The instrument is in the calibration cycle and has been self-calibrated.
“Typical” and “nominal” for this product are defined as follows
• Typical: Refers to the performance of the product under certain conditions.
• Nominal: Refers to the approximate value under product application process.
Frequency
Frequency Range
XSA805 (TG)
9.000 kHz to 500.009 MHz
XSA810 (TG)
9.000 kHz to 1.000009 GHz
Frequency Resolution
1 Hz
Frequency Span
Span Range
0 Hz, 100 Hz to max frequency of instrument
Span Uncertainty
± span / (sweep points-1)
Internal Reference Frequency
Reference Frequency
10.000000 MHz
Reference Frequency
Accuracy
±[ (days since last calibrate × freq aging rate) + temperature sta initial accuracy]
Temperature stability
0°C to 50°C, reference to 25°C < 0.5 ppm
Aging rate
<1 ppm/year
Frequency Readout Accuracy
Marker frequency resolution
span / (number of sweep points - 1)
Marker frequency uncertainty
±(frequency indication × reference frequency ac 1% × span + 10% × resolution
ban marker frequency resolution)
Frequency Counter
Resolution
1 Hz, 10 Hz, 100 Hz, 1 kHz
Uncertainty
±(frequency indication × reference frequency ac counter resolution)
Frequency Span
Range
0 Hz, 100 Hz to maximum frequency of instrument
Uncertainty
± span / (number of sweep points - 1)
SSB Phase Noise (20°C to 30°C, fc=1 GHz)
Carrier Offset
10 kHz
< -80 dBc/Hz
100 kHz
< -100 dBc/Hz
1 MHz
< -115dBc/Hz
Residual FM (20°C to 30°C, RBW = VBW = 1 kHz)
Residual FM
< 50 Hz (nominal)
Bandwidth
Resolution Bandwidth(-3dB)
1 Hz to 1 MHz (1-3-5-10 steps by sequence)
RBW accuracy
< 5%, typical