Active parameter setting
Use the number keypad, knob, and
arrow keys to set the value.
28
Attenuation
Display input attenuation setting
AMPTD
→ [Attenuation]
29
Detector type
Display detector type
[Detector]
30
Reference level
Reference level
AMPTD
→ [Ref Level]
31
Amplitude Scale Type
Log (logarithmic) or Line (linear)
AMPTD
→ [Scale Type]
32
Amplitude Scale
Display amplitude scale
AMPTD
→ [Scale/Div]
4.7 Build-in Help
The built-in help provides information that refers to every function key and menu key on the front panel. Users can view this
help information if required.
1. How to acquire built-in help
Press Help; a prompt about how to obtain help information will be shown
2. Page up and down
If there is more than one page of information, you can read the complete information by using the directional keys.
3. Close the current help information
Press “Help” again to close help.
4. Acquire the menu help
A message about how to obtain help information will be shown, press the menu keys to get the corresponding help.
5. Acquire the help information of any function key
A message about how to obtain help information will be shown, press any function key to get the corresponding help.
4.8 Basic Measurement
Basic measurements include, input signal frequency and amplitude display, marked by a frequency marker. Follow these four
simple steps below to implement input signal measurement.
Basic:
1. Setting center frequency;
2. Setting span and resolution bandwidth;
3. Activate marker;
4. Setting amplitude;
For example, to measure a 100MHz -20dBM signal, you must turn on the spectrum analyzer and ensure it is warmed up for
30 minutes to ensure measurement accuracy.
1. Equipment connection:
Connect the output terminal of signal generator to the RF Input 50Ω terminal of spectrum analyzer. Set the parameters as
follows:
Frequency 100 MHz
Amplitude -20 dBm
2. Setting parameters:
1) Press [Preset] to restore the analyzer to its factory-defined state. The Spectrum analyzer will display the spectrum from
9kHz to the maximum span width. The signal generated will display as a vertical line at 100MHz. Refer to Figure 4-8