59
MTS_OW100_IFU-orthogold100-US-K182682_A
Table 3 Electromagnetic Interference
Guidance and manufacturer’s declaration – electromagnetic emissions
The OW100
®
is intended for use in the electromagnetic environment specified below. The customer or the user of
the OW100
®
should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment
– guidance
RF emissions CISPR
11
Group 1
Class A
The OW100
®
uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
Due to the characteristics of the device, which are characterized by
the emission, the usage of the device is allowed in industrial areas
and hospitals (CISPR11, class A)
When used in a living area ( acc. CISPR 11 class B is required) this
device may not provide adequate protection against radio services
The User may change the position or realign the device.
Test description
Standard-
specification
IEC 60601-1-
2:2014
Compliance level
IEC 60601-1-2:2014
Test result for OW100
®
Cable emissions
EN
55011:2009+A1
:2010
Class A 150kHz
–
30MHz
Compliance *
Radiated
emissions
EN
55011:2009+A1
:2010
Class A 30MHz
– 1GHz
Compliance *
Harmonic
emissions
EN 61000-3-
2:2014
Class A AC Input
≤16Amps
Compliance *
Voltage
fluctuations/
flicker emissions
EN 61000-3-
3:2013
Class A
AC Input ≤16Amps
Compliance *