Table 5 Guidance and manufacturer´s declaration
Guidance and manufacturer’s declaration – electromagnetic immunity
The OW100
®
is intended for use in the electromagnetic environment specified below. The customer or the user of
the OW100
®
should assure that it is used in such an environment.
IMMUNITY test IEC 60601-1-2
TEST LEVEL
Compliance
level
Electromagnetic environment
–
guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the OW100
®
,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
150kHz to
80MHz @
3Vrms; 6V in
ISM bands
between
150kHz-80MHz
80% AM at
1kHz
3 V / 6V
d
= 1,2√
P
Radiated RF
IEC 61000-4-3
See table 2
See table 2
d
= 1,2√
P
80 MHz to 800 MHz
d
= 2,3√
P
800 MHz to 2,7 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey,
a
should be less than the
compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked
with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the OW100
®
is used exceeds the applicable RF
compliance level above, the OW100
®
should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating theOW100
®
.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.