36
MTS_OW100_IFU-orthogold100-US-K182682_A
WARNING
The OW100
®
is intended for use by medical professionals only.
The OW100
®
may cause radio interference or interfere with the
operation of nearby equipment.
It may become necessary to take appropriate actions, such as re-
orientation or re-arranging the devices or the shielding.
CAUTION
Floor surface!
Floors should be wooden or concrete or be covered with ceramic
tiles. If the floor is covered with a synthetic material, the relative
humidity must be at least 30%.
NOTE
Power Interruptions!
Any break in the electrical supply can cause a "RESET" of the
OW100
®
. In this case the device needs a few seconds to restart.
If the break in electrical supply is longer the device will switch off
automatically. In this case the equipment must be restarted in
accordance with the operating instructions.
Both types of break in the electrical supply will cause the
equipment to go into a safe condition.
NOTE
The OW100
®
uses high frequency (= HF) technology for its
internal functions. Devices sensitive to electromagnetic
interference (cardiac pacemakers, etc.) should not be operated
close to the OW100
®
.
For details on HF interferences please note table 2, table 3 , table
4, table 5
CAUTION
Interferences!
In this environment, interferences are possible from devices that
carry the symbol for non-ionizating electromagnetic radiation.
CAUTION
Maintenance instructions!
Before switching
on the device, check if the regular necessary
maintenance work is done.
Switch on OW100
®
1. Connect the provided power cable to the power socket of the OW100
®
and to a
wall socket (Figure 7).
2. Press the power switch. A self test starts and one blue button on the touch wheel
will flash for a few seconds. Do not push the confirmation button at that time.