65
MTS_OW100_IFU-orthogold100-US-K182682_A
Error
Reason
Action
Applicator cannot be
unlocked/removed
Motor of applicator plug pulling
mechanism defective
Call service
Water cartridge cannot be
exchanged
Internal error
Call service
Leakage during connected
cartridge
Damaged seals
Insert new cartridge. If
problem continues, call
service.
Error Messages
Interference and/or failure states of operation of the device are displayed
with error messages including error code.
There are different categories of messages:
This textbox means fatal error. If not solved by pressing o.k., switch
the device off and on. If the error appears again, call service. Error messages
are specified by a code E and a 3-digit number.
This textbox means warning. If not solved by pressing o.k., switch
the device off and on.
Warning messages are specified by a code W and a 3-digit number.
This textbox means information. Thereby exist 2 categories of
information: one to be confirmed, and one shown short at the display. Info
messages are specified by a code i and a 3-digit number.