WARNING
Electromagnetic Interference
The device emits electromagnetic radiation when acoustic waves
are released. Devices sensitive to electromagnetic interference
should not be operated close to the device (e.g. cardiac
pacemakers, etc.).
If electromagnetic interference between the extracorporeal
acoustic wave device and nearby electronic equipment is
suspected (as evidenced by erratic behavior with either device), it
is recommended to increase the distance between the devices
until proper operation resumes. If it is necessary to operate an
electronic device in close proximity to the acoustic wave device
during treatment, both devices should be tested for proper
simultaneous operation prior to clinical use.
CAUTION
Operation environment!
The OW100
®
is designed for use in professional medical
institutions. Don´t use the device close to HF-surgery devices.
Don
’t use the device outside of the HF-shielded area of a MRI-
room.
NOTE
Usage of the OW100
®
in living areas
Due to the characteristics of the device, which are characterized
by the emission, the usage of the device is allowed in industrial
areas and hospitals (CISPR11, class A)
When used in a living area ( acc. CISPR 11 class B is required)
this device may not provide adequate protection against radio
services
The User may change the position or realign the device.
WARNING
Stacked devices!
The use of this device next to other devices or with other devices
in stacked form should be avoided as this may cause incorrect
operation. If use in the manner described above is nevertheless
necessary, this device and the other devices should be observed.
to make sure they are working properly.
WARNING
RF communication devices!
Portable RF communications equipment (such as radios including
their accessories such as antenna cables and external antennas)
should not be used closer than 30 cm (12 inches) away from the
manufacturer's designated parts and wiring of the OW100
®
.
Failure to do so may result in a reduction of performance of the
device.