10
MTS_OW100_IFU-orthogold100-US-K182682_A
Do not direct the acoustic waves on internal air-filled organs (especially lungs).
All other Contraindications mentioned in scientific literature
Do not use in areas where skin surface is broken or an open wound exist.
Never use this device on children, the unconscious, or anyone who cannot give verbal
consent or warnings about pain.
Do not direct the acoustic wave treatment on air- filled internal organs (especially
lungs). Do not use on the back or chest.
Do not direct the acoustic wave treatment on esophagus.
Do not direct the acoustic wave treatment on the ears.
Do not direct the acoustic wave treatment on the head.
Do not direct the acoustic wave treatment on the spinal column.
Do not use in the presence of unexplained pain.
Do not treat wounds, infected areas, skin tumors, open sores, or scars from
recent surgery.
2.3
Warnings
WARNING
A warning indicates a situation which, if not
avoided, could result in death or serious injury.
The OW100
®
may only be used by trained physicians (medical specialist) with
adequate medical and technical experience in acoustic wave treatments.
Use of this device may give rise to undesirable heart reactions. The patient must
be continuously observed during the treatment.
If a patient experiences cardiac arrhythmia during treatment or a kind of uneasy
feeling, acoustic wave delivery should be terminated.
If a patient reports significant or unexpected pain, immediately stop the treatment
and consult a physician.
Avoid possible bruising. Use caution when determining a patient’s sensitivity
level.
This device should not be used over swollen or inflamed areas or skin eruptions.
2.4
Cautions
CAUTION
A caution indicates a situation which, if not avoided, may result in minor
or moderate injury to the user or patient or damage to the OW100
®
equipment or other property.
Potential effects associated with acoustic wave treatments include those listed below:
There is a potential for discomfort when using the OW100
®
on bony surfaces. Be aware
of the potential for bruising.
Use of this device may be painful or cause bruising or cavitation.
Always protect patients and medical personnel with appropriate noise protection
measures, e.g. provide ear protectors.