Example: 4-20 mA signal outputs to control room
Starting position:
A plan wants to read the measurement values from the CO2 channel to their DCS
1
. The
SWG100 has only one IO module and three sample points. All outputs are free. The CO2
values from the second sample point should be logged in the control room.
Follow steps must be done to do this:
1.
Wiring the analog-output 1 with the DCS: Wiring the analog-output (AO1+/AO1-) with the
DCS. For this, use a wiring with a conductor cross section between
0,2…2,5 mm².
2.
In the next step output 1 must be configured at the SWG100. To do this, open the menu:
EXTRAS/ CONFIGURATE ANALOG OUTPUTS
. The following menu overview will
be appeared (see screenshot below). To configure the IO1, select the first record (channel
I/O 1/1 SP1 / CH4) in the list and open it.
1
DCS=Distributed Control System
DCS