MRU GmbH
SWG 100 User Manual
Page
111
of
155
Position 1: Acid-gas filter (#56795)
Required materials
2x Acid gas filter (#56795), contained in the service- set.
Required tools:
Needle-nose pliers.
Exchange interval:
Exchange necessary, if filter turns from purple to white.
Steps:
1.
Remove exhausted acid gas filter from the
clips.
2.
Pull the viton tubes from the exhausted acid
gas filter. If it is necessary, use needle- nose
pliers.
3.
Plug the viton tubes on the new acid gas
filters. Be sure, that the tubing is correct. (The
single tubes are signed with “+” and “
-
“, see
sketch on the right side.)
Steps:
Pull the viton-tubes from the filter unit. If it is necessary, pliers can be used to solve the tubes
from the filter unit.
Remove the exhausted filter unit from the clip.
Plug the viton tubes on the filter unit. Push the filter unit on the clip.
- Socket
+ Socket
Remove the
exhausted filter
unit.
Install a new filter
unit.