3.
Press
F2 = FW update
. The analyzer will start the update from the SD-card.
Update the pcb-mainboard (Firmware-
Update with filename “1106mb.fwb”)
1.
Open the path:
E
XTRA
/D
EVICE INFO
2.
Press
F3 = sub.syst
. to open the menu “
S
UB SYSTEMS INFO
”.
3.
Press
F2 = details
, to open the details from the mainboard. Be sure, that the device is
“Mainboard” to update the pcb
-mainboard.
4.
The Update will start from the SD-card, if a firmware with the filena
me “1106mb
.
fwb” is
at the SD-card.
Update the NDIR-bench (Firmware-
Update with filename “1106ndir.fwb”)
1.
Open the path:
E
XTRA
/D
EVICE INFO
2.
Press
F3 = sub.syst
. to open the menu “
S
UB SYSTEMS INFO
”.
3.
Press
F2 = details
, to open the details from the NDIR bench. Be sure, that the device
is “NDIR bench” to update the
NDIR-bench.
4.
The Update will start from the SD-card, if a fi
rmware with the filename “1106ndir.fwb” is
at the SD-card.
Or
F3
Or
F3