MRU GmbH
SWG 100 User Manual
Page
74
of
155
Connection of the external control via 4-20 mA input signals
The signal inputs built a 4-bit binary number:
I4
–
I3
–
I2
–
I1: 0-11 mA=open=0; 11/12-20 mA=closed=1.
Status of external
signal source
Status number
Description
I4 I3 I2
I1
-
-
0
0
0
0
0
Automatic sampling point switching
0
0
0
1
1
Analyzer is sampling the point SP1 (*1, *2)
0
0
1
0
2
Analyzer is sampling the point SP2 (*1, *2)
0
0
1
1
3
Analyzer is sampling the point SP3 (*1, *2)
0
1
0
0
4
Analyzer is sampling the point SP4 (*1, *2)
0
1
0
1
5
Analyzer is sampling the point SP5 (*1, *2)
0
1
1
0
6
Analyzer is sampling the point SP6 (*1, *2)
0
1
1
1
7
Analyzer is sampling the point SP7 (*1, *2)
1
0
0
0
8
Analyzer is sampling the point SP8 (*1, *2)
1
0
0
1
9
Analyzer is sampling the point SP9 (*1, *2)
1
0
1
0
10
Analyzer is sampling the point SP10 (*1, *2)
1
0
1
1
11
Analyzer is “stand
-
by” (*3)
1
1
0
0
12
Analyzer is “stand
-
by” (*3)
1
1
0
1
13
Analyzer is “stand
-
by” (*3)
1
1
1
0
14
Remote reset of all system alarms
1
1
1
1
15
Analyzer is “stand
-
by” (*3)
-
-
-
-
I4
+
I3
I2
I1
+
+
+
Summary of Contents for SWG100 CEM
Page 1: ...SWG100 User Manual MRU GmbH D 74172 Neckarsulm ...
Page 2: ...MRU GmbH SWG 100 User Manual MRU GmbH D 74172 Neckarsulm 2 155 ...
Page 102: ......
Page 142: ...MRU GmbH SWG 100 User Manual Page 142 of 155 13 10 Option NDIR bench for CO2 analysis 0 100 ...
Page 154: ...MRU GmbH SWG 100 User Manual Page 154 of 155 14 Declaration of conformity ...
Page 155: ...MRU GmbH SWG 100 User Manual Page 155 of 155 V1 00 EN ...