Software-side
•
Open the menu:
E
XTRAS
/A
UX INPUT
CONFIGURATION
(1).
•
The overview screen appears. At the overview-screen, the installed AUX-inputs are listed. Every IO
module has four AUX-inputs. At the default settings all AUX-inputs are deactivated (OFF at the
overview-screen).
•
To activate an AUX-input, push
F1
(=details). The different IO-inputs can be selected with the up/down
arrow keys.
•
After one AUX-input is selected, the measurement item must be activated. To do this, turn the
measurement item from OFF to definable.
•
The measurement item “definable” is an individually measurement configuration, where the user can
configure by himself.
The measurement item “definable” is an individual configuration channel. The menu “AUX input details” contains
some pre-configurated settings, like for temperature, or pressure sensors.
The structure of the menu “AUX input details”, for the channel “definable” is
given at the screenshot below.
•
Section (1):
Show the operator, which analog-input will be changed.
•
Meas. item (2):
The operator can select the sample point. This means, that
the analog-output signal from the choose sample point will be transmitted to
the process control system.
•
Item name (3):
At this point the operator selects the measurement item,
which should be transmitted. At the chart below, the typical measurements
items are counted. Basically, all measurement-channels, which can be
measured, can be selected at this point.
•
Meas. unit (4):
The dimension o
f the measurement item (example: %,ppm…)
•
Sample point (5):
This point shows, which unit the transmit signal have. This
point cannot be changed.
•
Minimum (4mA) (6):
Here the operator entering the equivalent measurement
value, for a current of 4 mA.
•
Maximum (20mA) (6):
Here the operator entering the equivalent
measurement value, for a current of 20 mA. The IO module will create the
linear relation.
•
OFF (6):
Leave the menu. Alternate push
F2
.
+
+
1
2
3
4
5
6
7
Summary of Contents for SWG100 CEM
Page 1: ...SWG100 User Manual MRU GmbH D 74172 Neckarsulm ...
Page 2: ...MRU GmbH SWG 100 User Manual MRU GmbH D 74172 Neckarsulm 2 155 ...
Page 102: ......
Page 142: ...MRU GmbH SWG 100 User Manual Page 142 of 155 13 10 Option NDIR bench for CO2 analysis 0 100 ...
Page 154: ...MRU GmbH SWG 100 User Manual Page 154 of 155 14 Declaration of conformity ...
Page 155: ...MRU GmbH SWG 100 User Manual Page 155 of 155 V1 00 EN ...