MRU GmbH
SWG 100 User Manual
Page
57
of
155
Software-settings
•
Open the menu:
Extras/Analog output configuration
(1).
•
The overview screen appears. This menu shows, how much analog-outputs are
available and how the analog outputs are occupied.
At the overview screen all analog-outputs are listed. The amount of the analog-outputs
is dependent from the amount of the installed IO-modules. Every IO module has 4
analog-outputs. If two IO modules are installed, the entire amount of analog-outputs is
eight. At the list, the user finds the information which analog signal is carry out at which
analog-output channel. The follow list-notation can be found (example):
Reference:
1.
The number of the IO module. 1 means, first IO module.
2.
The value of the analog-output. 3 means, the third analog-output.
3.
The sample point, from where the signal comes. SP1 means, the first sample point.
4.
The signal-name. CH4 means, that this analog-output transfers CH4 signal.
This example means: At the third channel of the first IO module, the CH4 signal from the first
sample point will be transferred.
At the sketch below, the menu-screen and the analog-outputs at the first, of two IO modules are
shown.
1
2
3
7
4
AO1
AO2
AO3
AO4
I/O 1/3 SP1 / CH4
1
2
3
4
AO1
AO2
AO3
AO4
Summary of Contents for SWG100 CEM
Page 1: ...SWG100 User Manual MRU GmbH D 74172 Neckarsulm ...
Page 2: ...MRU GmbH SWG 100 User Manual MRU GmbH D 74172 Neckarsulm 2 155 ...
Page 102: ......
Page 142: ...MRU GmbH SWG 100 User Manual Page 142 of 155 13 10 Option NDIR bench for CO2 analysis 0 100 ...
Page 154: ...MRU GmbH SWG 100 User Manual Page 154 of 155 14 Declaration of conformity ...
Page 155: ...MRU GmbH SWG 100 User Manual Page 155 of 155 V1 00 EN ...