Table 8.2
MODULATION TABLE
FOR LEADLESS REMOTE SENSING
(TYPE 1)
Row
VMOD
(Vdc)
Vos(MOD)
1
2
3
0.0
10.0
9999
0.00
0.20
0.00
Table 8.3
MODULATION TABLE FOR
EMULATION OF A PHOTOVOLTAIC ARRAY (TYPE 0)
Row
VMOD
(Vdc)
Mod(VMOD)
Vo
(Vdc)
Io
(Adc)
1
2
3
4
5
6
7
8
9
10
11
12
13
0.000
2.085
3.645
4.690
5.210
5.415
5.730
6.040
6.250
6.460
6.665
6.770
9999
1.000
0.952
0.905
0.857
0.809
0.762
0.714
0.666
0.619
0.571
0.476
0.000
0.000
105
100
95
90
85
80
75
70
65
60
50
0
0.0
10.0
17.5
22.5
25.0
26.0
27.5
29.0
30.0
31.0
32.0
32.5
Note: Set the program voltage to 105 V.
8.5
Battery Charger
Figure 8.4 illustrates a temperature compensated battery charger for applications with lead acid
batteries. Diode D1, placed between the power supply and battery, blocks current from flowing
from the battery to the power supply. This eliminates any loading on the battery when the power
supply is off, but more importantly, prevents the battery from charging the power supply’s output
capacitors. With deployment of diode D1, remote sensing should be applied across the battery
terminals to compensate for the diode drop. By setting the voltage and current to the bulk charge
voltage and maximum charge current, the power supply will initially charge the batteries in current
mode control and then automatically crossover to voltage mode control when the batteries reach
the desired set point. The power supply can be programmed for time dependent, sequential step
operation, see Section 3.1.1.5, to equalize and float charge the batteries after bulk charging.
116
Summary of Contents for DP SERIES II
Page 1: ...OPERATING AND SERVICE MANUAL DP SERIES II DC POWER SUPPLIES...
Page 2: ......
Page 3: ...MAGNA POWER ELECTRONICS INC 39 ROYAL ROAD FLEMINGTON NJ 08822 January 6 2010...
Page 4: ......
Page 6: ...improper use of the equipment and accessories ii...
Page 28: ...Figure 1 2 DP Series air cooled 10 kW to 30 kW package drawing with D Version front panel 15...
Page 47: ...Figure 3 7 Set remote sense Figure 3 8 Set internal control 34...
Page 48: ...Figure 3 9 Set external control Figure 3 10 Set rotary input 35...
Page 49: ...Figure 3 11 Set keypad input Figure 3 12 Set external program input 36...
Page 50: ...Figure 3 13 Set remote input Figure 3 14 Set external interlock 37...
Page 53: ...Figure 3 15 Calibration 40...
Page 61: ...Figure 3 24 Calibration 48...
Page 63: ...Figure 3 25 a Local and b remote sensing 50...
Page 64: ...Figure 3 26 External a resistive b voltage and current programming 51...
Page 111: ...98...
Page 113: ...Figure 5 1 Configuration setup Figure 5 2 GPIB communications setup 100...
Page 118: ...Figure 5 7 Register Panel Figure 5 8 Calibration Panel 105...
Page 119: ...Figure 5 9 Firmware Panel Figure 5 10 Modulation Panel for Version D models only 106...
Page 121: ...Figure 6 1 Block diagram 108...
Page 139: ...Figure B 1 Information Panel Figure B 2 Configure Panel 126...
Page 140: ...Figure B 3 Reboot in Progress Panel Figure B 4 Web Control Panel 127...
Page 145: ...Figure D 1 RS485 a 2 wire and b 4 wire connections 132...