SENSOR DATA
Electrical Data
Supply:
20 VDC, 20 mA
RS485:
Half-duplex
Baud rate/device address:
19200 bps, 1
Power consumption:
0.4 W
Mechanical Data
Sensor Measuring Surface:
POM or Ceramic
Sensor Housing Material:
304 SS (1.4307)
Protection class:
IP 67 according to EN 60529
Sensor Weight (approx):
2.31 lb. (1,050 g)
Sensor Dimensions:
4.57” Dia. x 2.02” H (116 mm x 51.5 mm)
Response time (approx):
1 sec
Connecting cable:
Shielded cable, 4-wire, min.: 0.0008 in
2
(0.5 mm²)
Wire length:
max. 3280 ft. (1,000 m)
Limit Values
Pressure resistance:
max. 6 bar, short-term 10 bar
Working temperature:
+32 to +176° F (0° C to +80° C) in the sensor
Process temperature:
+14 to +194° F (-10° C to +90° C) at the measuring area,
with separation layer or cooling up to 284° F (140° C)
Storage temperature:
+14 to +176° F (-10° C to 80° C)
TRANSMITTER DATA
Moisture Measuring
Measured value:
0-85% moisture or 15-100% dry substance (TR)
Indicator:
Percentage value with max. 3 decimal places
Accuracy:
max. 0.2% (depending on the product)
Average value:
0-999 seconds
Filter value:
0-999.9 seconds
Temperature coefficient
Maximal:
<0.05 %/K
Typical:
<0.03 %/K
Product memory:
max. 24 product characteristics
General
Galvanic separation
In-/Output/Supply:
300 Veff (rated insulation voltage, overvoltage II, Contamination
level2, secure division according to EN 61010, EN 50178)
Electrical connection
Terminal 1 to Terminal 24:
Screw terminals
Terminal 1 to Terminal 5:
TBUS-connection with 0.004 in
2
(2.5 mm²)
Transmitter Housing
Material:
PBT
Protection class:
IP 20
Combustibility class:
VO to UL
Transmitter Dimensions (LxWxH): 4.53” x 0.89” x 3.94” without clamps (115 mm x 22.5 mm x 100 mm)
Transmitter Weight (approx): 0.55 lb. (250 g)
Form of construction:
Terminal case for assembling mounting rail
Mounting/Installation:
random
Ambient conditions
Process temperature:
+14 to +140° F (-10° C ... +60° C)
Storage/ transport:
+14 to +158° F (-10° C ... +70° C)
Process humidity while
operation:
10% … 95% humidity without condensation
26