15
On-Line display
Use the function Online display to call up the current real time measured values from the
Humi
Core
™
Pro
sensor (See Figure 23). After selection of the button ‘Online display’ the following values can be taken
from the screen:
Product:
Name of the product
Measured value: Current measured value
Temperature:
Current sample temperature
Raw value:
Current measured raw value
Output:
Current mA value of variable
A sensor or transfer error is indicated in the online
screen with
Sensor fault
(Figure 24).
The mA value for the analog output jumps to the
previously parameterized alarm value.
All other measured values are frozen when there is
an error and thus do not change any more.
Error elimination:
Check cable and terminal connections and possi-
bly restart the system.
PC-Settings
Serves the purpose of making port settings and selecting the menu language (see ‘Connect configuration
program’ in Section 6.3).
Figure 23
Figure 24