0112-0133 Rev B
139
8
Troubleshooting
The following error messages appear in the Status window. For more information,
Table 8-1
Troubleshooting table
Error
Code
Error Message
Description
Solution
200
Error: A problem was
encountered during
Reset.
Homing routine was
not able to find a state
change on the home
flag, sensor/flag
problem likely source.
Press the Power/Reset
button.
201
Error: Call Molecular
Devices Technical
Support at +1-800-
635-5577.
Timer resource was
not available,
apparently not used in
code.
Call Molecular
Devices Technical
S1-800-635-
5577.
202
Error: A problem
occurred during a
motion.
Stall occurred during
motion, encoder and
step counts exceeded
maximum threshold.
Check for object
blocking movement
and then re-calibrate.
203
Error: A problem
occurred with the X
axis.
X Axis motion did not
cause state change on
edge detect switch —
poorly gripped plate,
plate not present.
Re-calibrate the
instrument.
204
Error: A problem was
encountered during
Reset.
Can't move axis if it
hasn't been homed.
Press the Power/Reset
button.
205
Error: A problem
occurred during a
motion.
Can't move axis if it's
already in motion.
Press the Power/Reset
button.
206
Error: A Problem
occurred with the
EPROM.
Generic EPROM error. Call Molecular
Devices Technical
S1-800-635-
5577.
207
Error: A problem
occurred during a
motion.
Carriage drop off
exceeds maximum
travel of X Axis.
Re-calibrate the
instrument.
Summary of Contents for StakMax
Page 1: ...StakMax Plate Handling System User Guide 0112 0133 Rev B May 2010...
Page 6: ...Contents 6 0112 0133 Rev B General Specifications 141 Diagrams 142...
Page 10: ...Safety 10 0112 0133 Rev B...
Page 14: ...Introduction 14 0112 0133 Rev B...
Page 80: ...Installation and Setup of SpectraMax L instrument 80 0112 0133 Rev B...
Page 116: ...Operation 116 0112 0133 Rev B...
Page 146: ...System Diagrams and Specifications 146 0112 0133 Rev B...