Operation
110
0112-0133 Rev B
Emergency Stop
The
Emergency Stop
button immediately shuts down the entire system,
including the StakMax system and the reader. After you click the
Emergency Stop
button, clear any object that is in the way, remove any
plates that are in the drawer, and then click the
Reset
button in the main
StakMax
window.
Load Plate
The
Load Plate
button manually moves one plate from the input stack to
the reader. Instead of running a complete calibration, you can use this
command to perform a “quick check” to make sure that the plate is
aligned and moving properly between the input stack and the reader.
Return Plate
The
Return Plate
button manually moves one plate from the reader to
the output stack. You can use this command to clear a plate that was left
in the reader or to help determine if a component is out of sync.
Restack
The
Restack
button moves plates from the output stack to the input stack.
Specify the number of plates to restack. To move all of the plates, leave
the Number of Plates field blank or type 0 (zero). See
.
Figure 5-21
The Restack Plates window.
Summary of Contents for StakMax
Page 1: ...StakMax Plate Handling System User Guide 0112 0133 Rev B May 2010...
Page 6: ...Contents 6 0112 0133 Rev B General Specifications 141 Diagrams 142...
Page 10: ...Safety 10 0112 0133 Rev B...
Page 14: ...Introduction 14 0112 0133 Rev B...
Page 80: ...Installation and Setup of SpectraMax L instrument 80 0112 0133 Rev B...
Page 116: ...Operation 116 0112 0133 Rev B...
Page 146: ...System Diagrams and Specifications 146 0112 0133 Rev B...