B.1 TCP/IP
53
approach will work via
USB
. Type a command into the command line win-
dow, and hit
Enter
to send that command to the device (see appendix D);
the response will appear in the box below the command line window. Some
commands are not permitted because they require complex data flow.
B.1.2
Configuration in alternate subnet
Another approach is to plug the device into a network which is in the
same address space, then send the appropriate commands by
telnet
, and
reconnect in the desired address space. For example, if the
MWM
default
address is 10.1.1.180, and you want the device to work in the 192.168.1.xxx
space:
1. Use a network patch cable to connect the
MWM
directly to a computer
ethernet socket, not via a network switch or hub.
2. Configure the computer network settings to have an address in the
same address space as the
MWM
; for this example, you can set the
computer
IP
address to 10.1.1.189.
Summary of Contents for MWM