30
Chapter 4. MOGWAVE
The row can change, due to thermal drift or if the fibre is changed, in
particular if changing from FC/PC to FC/APC or vice versa. The row is
tracked automatically but the row must initially be close, to within about
10 pixels of the centre of the spectral peak.
The row can be set via the up/down arrows, typed in numerically, or using
the mouse inside the focus window as described above. Again, note that
the initial starting value of the row is programmed in the device
EEPROM
but can easily be updated as described above.
Auto-row
The automatic tracking of the row for the main spectral peak can be
enabled or disabled via the check-box.
Summary of Contents for MWM
Page 1: ...MWM wavemeter Revision 3 19 mogwave 1 4 15 Firmware 0 6 9 ...
Page 4: ...ii ...
Page 7: ...Contents v G Ugrading firmware 83 References 84 ...
Page 8: ...vi Contents ...
Page 54: ...46 Chapter 6 Calibration ...
Page 58: ...50 Appendix A Specifications ...
Page 66: ...58 Appendix B Communications ...
Page 70: ...62 Appendix C Programming ...
Page 86: ...78 Appendix E mogwave configuration ...
Page 92: ......
Page 94: ...86 ...
Page 95: ......