6-9
6-9
6-9
6-9
6-9
Creating Programs
The gradient screen will appear:
Enter: CUSTOM1
Lower Temp
°
C:
Enter the lower limit temperature (for the purposes of this example, 50
°
),
then press «
Proceed
». The upper temperature screen will appear:
Enter: CUSTOM1
50.0
°
Upper Temp
°
C:
Enter the upper temperature (for the purposes of this example, 70
°
), then
press «
Proceed
». (Use integers only; decimals are not accepted).
The next screen requires you to enter a hold time for the temperature
gradient step:
Enter: CUSTOM1
2= 50.0
°
to 70.0
°
Time:
Enter the hold time in the form of min:sec. Press «
Proceed
». A confirma-
tion screen will appear:
Enter: CUSTOM1
2= 50.0
°
to 70.0
°
for 0:30
OK? Yes No Option
Select “Yes” to enter the step into memory and proceed to the next step;
select “No” to reject or edit the current step. To preview, select
Option
,
then
Preview
.
Preview
gives the predicted temperatures for each of the
wells along the sample block’s long axis. (The lines do not all appear in
the display; use the «
Select
» keys to scroll up or down.)
Gradient Preview:
Column 1— 50.0
°
left
Column 2— 51.8
°
Column 3— 53.6
°
Summary of Contents for PTC-0200
Page 6: ...Part I The DNA Engine ...
Page 7: ......
Page 11: ......
Page 23: ......
Page 76: ......
Page 92: ......
Page 96: ......
Page 105: ...Part II The DNA Engine Tetrad ...
Page 106: ......
Page 117: ...Part III Accessories ...
Page 118: ......
Page 125: ...13 7 13 7 13 7 13 7 13 7 The Remote Alpha Dock System Figure 13 5 Flush mounting template ...
Page 129: ......
Page 131: ......
Page 133: ......