Gradient Specifications (96V Alpha module only)
Accuracy:
+ 0.4
°
C of programmed target at end columns, 30
seconds after the timer starts for the gradient step,
NIST–traceable
Column uniformity:
+ 0.4
°
C , well–to–well within column, within 30 sec-
onds of reaching target temperature
Calculator accuracy:
+ 0.4
°
C of actual well temperature
Lowest programmable
30
°
C
temperature
Highest programmable
105
°
C
temperature
Temperature differential
1–24
°
C
range