DNA Engine & DNA Engine Tetrad Operations Manual
Figure 13-3
Attachment of power and data cables
•
Remote Docks with mounted Alpha units may be stacked vertically
as long as a minimum bottom clearance is maintained that would be
no less than that resulting from the unit being placed on a solid hori-
zontal platform. A minimum top clearance is also required to allow
access to and operation of the Alpha unit lid.
•
Remote Docks can be flush-mounted (i.e., with the feet removed) to
facilitate robotic operation, as long as the airway beneath the unit is
equivalent to the airway the unit would have with the feet attached.
Usually a hole will need to be cut to allow air to flow to the the
cooling fan. Figure 13-5 is a template for flush-mounting the Remote
Dock.
Power Supply
•
The Alpha unit mounted in each Remote Dock is powered from the
PTC-200 or PTC-225 base.
•
The Remote Dock’s fan is powered externally, and a power supply is
provided that requires power from 90-250 VAC and 47 to 63 Hz, with
a grounded outlet.
Air Supply
•
Alpha units being operated in the remote configuration have no op-
erating constraints that do not also apply to normal operations in the
PTC-200 or PTC-225.
Dock connector
Remote dock
Summary of Contents for PTC-0200
Page 6: ...Part I The DNA Engine ...
Page 7: ......
Page 11: ......
Page 23: ......
Page 76: ......
Page 92: ......
Page 96: ......
Page 105: ...Part II The DNA Engine Tetrad ...
Page 106: ......
Page 117: ...Part III Accessories ...
Page 118: ......
Page 125: ...13 7 13 7 13 7 13 7 13 7 The Remote Alpha Dock System Figure 13 5 Flush mounting template ...
Page 129: ......
Page 131: ......
Page 133: ......