32
4
Operation
Trendelenburg
1. Press
and
simultaneously.
The
tabletop will move to the head up position.
n
Reverse Trendelenburg
(Head up)
n
Trendelenburg (Head down)
Head up
NOTE
- The maximum angle of
head up or head down is
25 degrees from the level
position.
- When the tabletop in a
Trendelenburg or reverse-
Trendelenburg position is tilted
to the opposite side, it will stop
in the level position. Once it
stops, release the switch and
press it again; the tabletop will
continue tilting.
25°
25°
25°
25°
WARNING
Make sure the patient is
secured on the table, for
example with a patient safety
strap. Otherwise, the patient
may fall of the table.
1. Press
and
simultaneously.
The tabletop will move to the head down
position.
Head down
SEAT DOWN
SEAT UP
SEAT DOWN
SEAT UP
Summary of Contents for MOS-1302B
Page 4: ......
Page 5: ...1 1 Introduction 1 Introduction ...
Page 6: ...2 1 Introduction ...
Page 9: ...5 2 Before use 2 Before use ...
Page 18: ...14 2 Before use ...
Page 19: ...15 3 Component identification 3 Component identification ...
Page 22: ...18 3 Component identification ...
Page 48: ...44 4 Operation ...
Page 49: ...45 5 Using the leg traction device 5 Using the leg traction device ...
Page 66: ...62 5 Using the leg traction device ...
Page 67: ...63 6 Maintenance and inspection 6 Maintenance and inspection ...
Page 70: ...66 6 Maintenance and inspection ...
Page 71: ...67 7 Specifications 7 Specifications ...
Page 74: ...70 7 Specifications Appearance Operating Table Extension arm right left ...
Page 75: ...71 8 When the control unit can not be used 8 When the control unit can not be used ...
Page 77: ...73 9 Troubleshooting 9 Troubleshooting ...
Page 80: ...9 Troubleshooting 76 ...
Page 81: ...77 APP Electromagnetic compatibility App Electromagnetic compatibility ...