background image

MIROxi

 

User Manual cod. 980265  

Rev 1.6   

Page 6 of 32 

Use only MIR sensors supplied with or specifically intended for use with the 

MIROxi

. The use of sensors not 

intended for use with the 

MIROxi

 may cause inaccurate readings. 

Oximetry measurements may be inaccurate in the presence of high ambiental light. Shield the sensor area 

(with a surgical towel, for example) if necessary.  

WARNING

Dyes introduced  into the bloodstream, such as methylene blue, indocyanine green, indigo carmine, patent 

blue V (PBV), and fluorescein may adversely affect the accuracy of the oximetry reading.  

 

Any condition that restricts blood flow, such as the  use of a  blood  pressure cuff or a device for  systemic 

vascular resistance, may cause an inability to determine accurate pulse rate and SpO2 readings. 

Remove fingernail polish and/or false fingernails before applying SpO2 sensors. Both may cause inaccurate 

oximetry measurement. 

 

Significant  levels  of  dysfunctional  hemoglobins,  such  as  carboxyhemoglobin  or  methemoglobin,  may 

adversely affect the accuracy of the oximetry measurement. 

 

Optical cross-talk can occur when two or more sensors are placed in close proximity. Optical cross-talk may 

adversely affect the accuracy of the oximetry readings. The danger can be eliminated by covering each site 

with opaque material. 

Obstructions  or  dirt  on  the  sensor’s  emitter  and/or  detector  may  cause  a  sensor  failure  or  inaccurate 

readings. Make sure there are no obstructions and the sensor is clean. 

 

Autoclaving, ethylene oxide sterilizing, may cause sensor damage. Do not attempt to sterilize the sensor. 

 

Unplug the sensor from the 

MIROxi

 before cleaning or disinfecting to prevent damaging sensor or device, 

and to prevent safety hazards for the user. 

 

1.2.3 

Device 

WARNING

The  maintenance  operations  detailed  in  this  Manual  must  be  fully  and  accurately  carried  out.  If  these 

instructions are not followed this can cause measurement errors and/or an incorrect test interpretation.  

Any  modifications,  adjustments,  repairs  or  reconfiguration  must  be  made  by  the  manufacturer  or  by 

personnel  authorised  by  the  manufacturer.  Never  attempt  to  make  a  repair  oneself.  The  set-up  of 

configurable  parameters  should  only  be  made  by  qualified  personnel.  However,  an  incorrect  set-up  of the 

parameters does not put the patient at risk. 

High-frequency  emissions  from  other  electronic  devices  may  interfere  with  the  correct  operation  of  the 

instrument.  For  this  reason,  certain  minimum  clearances  (a  few  metres)  should  be  observed  when  high-

frequency appliances such as a TV, radio, portable phone, etc. and other electronic units are operated at the 

same time in the same room.  

The instrument may give inaccurate readings if operated in the presence of strong electromagnetic sources, 

such as electrosurgical equipment, or in the presence of computed tomography (CT) equipment. 

 

Do not use the instrument in the presence of magnetic resonance imaging (MRI) equipment. MRI equipment 

may cause an induced current to the oximetry sensor, resulting in patient injury. 

If the PC and/or the printer connected to 

MIROxi

 come into contact with an area containing patient data, ref. 

directive EN 60601-1-1, it is necessary that they conform to the directive EN 60601-1. 

For the recycling of the 

MIROxi

, the accessories, any plastic consumable materials  as well as the battery, 

use  only  the  appropriate  containers  or  return  all  such  parts  to  the  dealer  or  to  a  recycling  centre.  All 

applicable local regulations must be followed.  

If  any  of  these  rules  are  not  followed  then  MIR  will  decline  all  responsibility  for  any  direct  or  indirect 

damages, however caused. 

Use only the battery type indicated in the § Technical specifications. 

Remove the battery from the device if the machine is not used for a long period (several months). 

The instrument may be powered through a PC by a USB cable. By this means, the device works both on line 

with the PC, or individually powered by the PC. 

Keep the device out of reach of children and of any person with mental handicap. 

 

1.3 

Unforeseen errors 

 

In case the device internal memory data are damaged, when the device is switched on, the following message appears: 

Error on RAM memory 

Recovery data 

Please wait 

 

If data are successfully recovered then the device completes the standard switch on process, otherwise please contact 

an authorised technical assistance point or the manufacturer. 

Summary of Contents for MIROxi

Page 1: ...MIROxi User Manual cod 980265 Rev 1 6 Page 1 of 32 MIROxi User Manual User Manual Rev 1 6 Issued on 06 10 2008 Approved on 06 10 2008...

Page 2: ...hardware will be recognised by the PC The device can then be used with the winspiroPRO software Keep the original packaging In the event that your device requires service then always use the original...

Page 3: ...evel 13 2 3 Information 13 2 4 Initial Set up 13 2 5 Patient Data 15 2 6 Displaying data in memory 16 2 7 Online operating mode connected to a PC 16 2 8 Oximetry testing 16 2 8 1 Walk Test 6MWT 18 2 8...

Page 4: ...her own health The procedures for using the device at home are described according to the type of test to be made the display will show all instructions messages suggestions etc step by step which all...

Page 5: ...tery is discharged Contact Technical Service to have this battery changed In the event of any incident or accident of any kind resulting from the use of the device the user is required to inform the m...

Page 6: ...manufacturer Never attempt to make a repair oneself The set up of configurable parameters should only be made by qualified personnel However an incorrect set up of the parameters does not put the pati...

Page 7: ...1 4 2 CE mark for medical devices 0476 This product is certified to conform to the Class II requirements of the 93 42 EEC medical device directive 1 4 3 Electrical safety symbol In accordance with the...

Page 8: ...his equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by...

Page 9: ...BPM hh mm ss T 120 BPM Time passed with pulse frequency 120 BPM hh mm ss N Events 40 BPM Bradycardia events during the entire period of analysis N Events 120 BPM Tachycardia events during the entire a...

Page 10: ...f the method for calculating the area below the SpO2 baseline curve Parameters requested for six minute walk test analysis SYMBOL DESCRIPTION Units Dyspnea Borg Baseline Grade of dyspnea before walkin...

Page 11: ...brane keyboard with 6 keys Interface RS232 USB Bluetooth Duration of the CR2032 3V lithium battery memory backup Circa 10 years under normal conditions of use and storage Power supply 4 x AAA batterie...

Page 12: ...et up Service Menu To manage or enter new patient data from the main screen To enter new patient data To modify patient data To carry out an oximetry test To access previous tests To access transmissi...

Page 13: ...i x y device name and software revision number Power means USB or battery symbol Current date and time icon Access to the Service Menu SpO2 BPM icon Press OK to go to the second display Press ESC to g...

Page 14: ...Oxi will start to search for Bluetooth devices in the area once one or more devices are found the screen will display profiles press OK to set the device as printer or phone select one of the two opti...

Page 15: ...y To enter new patient data From the Patient Data Management menu press to access the data input screen First screen name and surname Use and to select required letters confirm or delete using or To i...

Page 16: ...ut searching Use from the data screen icon to print out data through a Bluetooth connection automatic use to display all test data of relevant file on memory use and to scroll through data press OK to...

Page 17: ...DESCRIPTION Walk test 6MWT Sleep oximetry Oximetry SpO2 BPM WARNING In order not to compromise the reproducibility of the measurements and the integrity of the sensor avoid twisting the sensor cable...

Page 18: ...o pass to the following walking phase The number of bars I symbol on the right upper of the screen is proportional to the quality of the oximetry signal the higher the quality of the signal the more b...

Page 19: ...i will go on standby i e it stops beeping and the display turns off The led signal remains on To control the correct functioning while on standby press then after 1 minute MIROxi will return to standb...

Page 20: ...se of this sensor is indicated for patients weighing more than 30 Kg and contraindicated for patients with allergy to adhesive tape WARNING The materials used for manufacturing the sensor are NATURAL...

Page 21: ...rious tests to the doctor s PC follow this procedure Dial the doctor s phone number for example the doctor s office telemedicine services or any other structure directly on the telephone in use Place...

Page 22: ...or s office telemedicine service etc To enter the telephone number see the main menu when the machine is turned on Refer to Paragraph 2 4 A device must also be set up for the connection refer to Parag...

Page 23: ...ortant prior to initiating the following procedure to know the version of the operating system installed on the PC used for the connection from control panel click on System where the type of operatin...

Page 24: ...nsor has alignment pins Place the sensor with the alignment pins facing the adhesive part of the tape and align the pins to the holes on the tape Push the sensor to insert the pins into the holes of t...

Page 25: ...of oximetry testing the test data is incorrect Clean the sensor and then control it otherwise use a new one All data in memory lost due to an unforeseeable event All data in memory has been deleted Co...

Page 26: ...97 Notified Body CERMET No 0476 Certificate No MED 9826 MIR srl Medical International Research declares that the Device subject of this declaration together with its standard accessories conforms to...

Page 27: ...utilised differently from the use described in the User s Manual If any alteration adjustment modification or repair has been carried out by personnel not authorised by the manufacturer If the fault...

Page 28: ...MIROxi User Manual cod 980265 Rev 1 6 Page 28 of 32 ANNEX 1 OXYMETRY TEST REPORT EXAMPLES...

Page 29: ...MIROxi User Manual cod 980265 Rev 1 6 Page 29 of 32...

Page 30: ...MIROxi User Manual cod 980265 Rev 1 6 Page 30 of 32 MIROxi...

Page 31: ...4 1 kV for input output lines Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode Not Applicable Mains power...

Page 32: ...een portable and mobile RF communications equipment and the MIROxi The MIROxi is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the...

Reviews: