MIROxi

 

User Manual cod. 980265  

Rev 1.6   

Page 23 of 32 

Printing  of  data  from  the  patient  management  function  is  enabled  only  if  the  printer  has  a  Bluetooth 

connection; alternatively a USB key can be installed on the printer in order to enable a Bluetooth connection. 

 

The  Bluetooth  system  enables 

MIROxi

  to  transfer  test  data  directly  to  a  Bluetooth  enabled  printer.  The  steps  are  as 

follows:  

 

  From the main screen select the test to be printed with  . 

  On the next screen press   (

 icon). 

  By selecting oximetry tests, the most recent test made by that patient will be printed. 

 

Tests stored in memory can also be printed. Use the search method as described in Paragraph 2.6 to print out relevant 

tests. 

 

  On the relevant test screen press  . 

 

MIROxi

 will carry out the connection. 

  At  the  end  of  transmission 

MIROxi

  will  show  “CONNECTION  COMPLETED”,  and  return  automatically  to  the  main 

screen. 

 

To interrupt the Bluetooth connection during transmission press ESC to return to the main screen. 

Where no printer has been set up, a message will appear to search for devices. After the device has been set up it will 

automatically be enabled for printing data. 

When searching for Bluetooth enabled devices, 

MIROxi

 will check the address of that device and where a previously 

registered device has changed name, it will be automatically updated. 

 

3.4  Connection to a PC through USB port 
WARNING

 

Before connecting 

MIROxi

 by USB to a PC, winspiroPro must be installed to interface with the device. 

It  is  important  prior  to  initiating  the  following  procedure  to  know  the  version  of  the  operating  system 

installed  on  the  PC  used  for  the  connection  (from  control  panel  click  on  “System”,  where  the  type  of 

operating system installed on the PC can be checked). 

If winspiroPRO is already installed on the PC then a new installation is not required. 

 

To make the connection, insert the mini USB connector supplied with 

MIROxi

 

as shown in the picture and attach the other connector to the USB port of the 

PC. 

When making the initial connection, the PC will (depending on the version of the 

operating system) either make an automatic driver installation or request some 

information.  To  avoid  errors  in  this  phase  please  read  the  winspiroPRO  User 

Manual very carefully. 

To check the connection between the device and the PC ensure that the led on 

the device is illuminated.

 

 

 

3.5  Connection to a PC through the RS232 

 

MIROxi

  can  also  be connected to  a PC through the RS232 serial port. 

This  leaves  the  USB  port  free  and  the  device  is  run  directly  by  the 

software  (as  a  USB  connection).  The  picture  shows  the  RS  232 

connector attached to 

MIROxi

For the correct management of the device see the online manual of the 

software.

 

 

 

3.6  Upgrade Internal software 

MIROxi

  software  can  be  upgraded  when  connected  to  a  PC  (USB  or  RS232)

.

  Upgrades  can  be  downloaded  by 

registering  on  www.spirometry.com.  For  further  information  on  upgrading  software  see  the  “winspiroPro”  software 

manual. 

 

4  MAINTENANCE 

 

MIROxi

 is an instrument that requires very little maintenance. The operations to perform periodically are:  

 

  Cleaning of the oximetry sensor (for reusable sensors). 

 

Changing the adhesive tape of the oximetry wrap sensor.

 

Summary of Contents for MIROxi

Page 1: ...MIROxi User Manual cod 980265 Rev 1 6 Page 1 of 32 MIROxi User Manual User Manual Rev 1 6 Issued on 06 10 2008 Approved on 06 10 2008...

Page 2: ...hardware will be recognised by the PC The device can then be used with the winspiroPRO software Keep the original packaging In the event that your device requires service then always use the original...

Page 3: ...evel 13 2 3 Information 13 2 4 Initial Set up 13 2 5 Patient Data 15 2 6 Displaying data in memory 16 2 7 Online operating mode connected to a PC 16 2 8 Oximetry testing 16 2 8 1 Walk Test 6MWT 18 2 8...

Page 4: ...her own health The procedures for using the device at home are described according to the type of test to be made the display will show all instructions messages suggestions etc step by step which all...

Page 5: ...tery is discharged Contact Technical Service to have this battery changed In the event of any incident or accident of any kind resulting from the use of the device the user is required to inform the m...

Page 6: ...manufacturer Never attempt to make a repair oneself The set up of configurable parameters should only be made by qualified personnel However an incorrect set up of the parameters does not put the pati...

Page 7: ...1 4 2 CE mark for medical devices 0476 This product is certified to conform to the Class II requirements of the 93 42 EEC medical device directive 1 4 3 Electrical safety symbol In accordance with the...

Page 8: ...his equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by...

Page 9: ...BPM hh mm ss T 120 BPM Time passed with pulse frequency 120 BPM hh mm ss N Events 40 BPM Bradycardia events during the entire period of analysis N Events 120 BPM Tachycardia events during the entire a...

Page 10: ...f the method for calculating the area below the SpO2 baseline curve Parameters requested for six minute walk test analysis SYMBOL DESCRIPTION Units Dyspnea Borg Baseline Grade of dyspnea before walkin...

Page 11: ...brane keyboard with 6 keys Interface RS232 USB Bluetooth Duration of the CR2032 3V lithium battery memory backup Circa 10 years under normal conditions of use and storage Power supply 4 x AAA batterie...

Page 12: ...et up Service Menu To manage or enter new patient data from the main screen To enter new patient data To modify patient data To carry out an oximetry test To access previous tests To access transmissi...

Page 13: ...i x y device name and software revision number Power means USB or battery symbol Current date and time icon Access to the Service Menu SpO2 BPM icon Press OK to go to the second display Press ESC to g...

Page 14: ...Oxi will start to search for Bluetooth devices in the area once one or more devices are found the screen will display profiles press OK to set the device as printer or phone select one of the two opti...

Page 15: ...y To enter new patient data From the Patient Data Management menu press to access the data input screen First screen name and surname Use and to select required letters confirm or delete using or To i...

Page 16: ...ut searching Use from the data screen icon to print out data through a Bluetooth connection automatic use to display all test data of relevant file on memory use and to scroll through data press OK to...

Page 17: ...DESCRIPTION Walk test 6MWT Sleep oximetry Oximetry SpO2 BPM WARNING In order not to compromise the reproducibility of the measurements and the integrity of the sensor avoid twisting the sensor cable...

Page 18: ...o pass to the following walking phase The number of bars I symbol on the right upper of the screen is proportional to the quality of the oximetry signal the higher the quality of the signal the more b...

Page 19: ...i will go on standby i e it stops beeping and the display turns off The led signal remains on To control the correct functioning while on standby press then after 1 minute MIROxi will return to standb...

Page 20: ...se of this sensor is indicated for patients weighing more than 30 Kg and contraindicated for patients with allergy to adhesive tape WARNING The materials used for manufacturing the sensor are NATURAL...

Page 21: ...rious tests to the doctor s PC follow this procedure Dial the doctor s phone number for example the doctor s office telemedicine services or any other structure directly on the telephone in use Place...

Page 22: ...or s office telemedicine service etc To enter the telephone number see the main menu when the machine is turned on Refer to Paragraph 2 4 A device must also be set up for the connection refer to Parag...

Page 23: ...ortant prior to initiating the following procedure to know the version of the operating system installed on the PC used for the connection from control panel click on System where the type of operatin...

Page 24: ...nsor has alignment pins Place the sensor with the alignment pins facing the adhesive part of the tape and align the pins to the holes on the tape Push the sensor to insert the pins into the holes of t...

Page 25: ...of oximetry testing the test data is incorrect Clean the sensor and then control it otherwise use a new one All data in memory lost due to an unforeseeable event All data in memory has been deleted Co...

Page 26: ...97 Notified Body CERMET No 0476 Certificate No MED 9826 MIR srl Medical International Research declares that the Device subject of this declaration together with its standard accessories conforms to...

Page 27: ...utilised differently from the use described in the User s Manual If any alteration adjustment modification or repair has been carried out by personnel not authorised by the manufacturer If the fault...

Page 28: ...MIROxi User Manual cod 980265 Rev 1 6 Page 28 of 32 ANNEX 1 OXYMETRY TEST REPORT EXAMPLES...

Page 29: ...MIROxi User Manual cod 980265 Rev 1 6 Page 29 of 32...

Page 30: ...MIROxi User Manual cod 980265 Rev 1 6 Page 30 of 32 MIROxi...

Page 31: ...4 1 kV for input output lines Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode Not Applicable Mains power...

Page 32: ...een portable and mobile RF communications equipment and the MIROxi The MIROxi is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the...

Reviews: