MIROxi
User Manual cod. 980265
Rev 1.6
Page 25 of 32
5.1 Causes and Solutions
•
MIROxi does not switch on
Check that batteries are correctly inserted in the compartment on the back of the instrument. If they are correctly
positioned then replace them with new ones.
•
During operation the machine switches itself off and on again
Change the batteries.
•
At the end of oximetry testing the test data is incorrect
Clean the sensor and then control it; otherwise use a new one.
•
All data in memory lost due to an unforeseeable event
All data in memory has been deleted. Contact the technical service centre.
WARNING
Before contacting the centre for technical assistance, please download the archive from the device on to the
PC using the WinspiroPRO software. This procedure is necessary as the data could be lost whilst repairing
the device and could also be of a confidential nature and therefore subject to laws on data privacy.