9
(4) Connecting with load cell, Junction box B-304B for extension, and CSD-903
1) When you use CAB-502 (4 core cable)
・
When you use it with 4 wires method, be sure to make a short-circuited between A-F and between
C-G of the load cell terminal board. When terminal F and terminal G are used with opening status,
CSD-903 cannot make normal operation.
・
There is the possibility of the descent of the input voltage of the instrument at the time of total 30m
or more in length of CAB-502 (our standard cable) and also the possibility of outside of accuracy
due to the resistance of the cable.
B-304B internal terminal connection