97
10-11. Flow/Discharge brand setting target
The flow/discharge brand setting target is set by SQ function SQF-24.
SQF-26: Width for the automatic free fall compensation, SQF-37: Compensation flow time, and
SQF-38: Waiting time for judge after compensation flow can be set for each brand number selected
as the flow/discharge brand setting target.
SQF-24: Flow/Discharge brand setting target setting is not stored in EEPROM. Returns to default
when the power is turned off and back on again.
Setting range: 0 to 7
Default is set to [0].
10-12. Automatic free fall compensation
This is a function to guess the next Free Fall from a former actual Free Fall, and to set the value of
the Free Fall automatically when the measurement is operated continuously.
Set data of the next Free Fall uses the moving averaged value calculated from 4 times of actual
former Free Fall values.
10-12-1. Setting of automatic free fall compensation
The operation of Automatic free fall compensation is selected by SQ function SQF-25.
Default is set to [Invalid].
10-12-2. Width for the automatic free fall compensation
The width for the automatic free fall compensation is set by SQ function SQF-26.
The width for the automatic free fall compensation can be set for each brand number by setting
SQF-24: Flow/Discharge brand setting target.
Setting range: 00000 to 99999, Unit: 1D, 00000: OFF
Default is set to [00000].
10-12-3. Save of automatic free fall value
The operation of save of automatic free fall value is selected by SQ function SQF-27.
When save of automatic free fall value is set as [VALID], the value compensated by automatic free fall
compensation is saved setting value of free fall (SET2).
Default is set to [Invalid].
10-13. Judge condition
Judge condition at the time of finishing measuring is selected by SQ function SQF-30.
Judge condition at the time of finishing measuring is selected from [Timer is up], [Stable and timer is
up] or [Stable or timer is up].
Default is set to [Timer is up].
10-14. Waiting time for judge
Set the waiting time until executing judgment after D-FLOW is finished by SQ function SQF-31
Setting range: 0000-9999, Unit: 0.01s, 0000: OFF
Default is set to [0000].
Summary of Contents for CSD-903-EX
Page 1: ...DIGITAL INDICATOR CSD 903 EX Instruction Manual EN294 1492 H ...
Page 2: ...IV ...
Page 91: ...75 ...
Page 93: ...77 ...
Page 95: ...79 ...
Page 97: ...81 ...
Page 199: ...183 Inform us about the contents of failure and situation at site in details ...
Page 204: ...188 Inform us about the contents of failure and situation at site in details ...
Page 206: ...190 Inform us about the contents of failure and situation at site in details ...
Page 209: ...193 Inform us about the contents of failure and situation at site in details ...
Page 224: ... ...
Page 225: ......