107
12-7. Confirmation of load cell output voltage
In the confirmation of the load cell output voltage, the weight put on a present load cell is converted
into mV/V.
For example, when an uncertain load cell is used, please execute the calibration based on the load
cell output value with the actual weight for the tare weight and weighing capacity.
By pressing the
key from
display, the
display indicates the present output voltage value of the load cell
and blinks.
: Quit from check mode.
The displayed load cell output voltage value is a reference value. The display accuracy is about 0.5%.
12-8. Confirmation of BCD output (When BCD output is applied.)
By pressing the
key from
display,
is displayed, and 100 digit blinks.
At this time, the same value as the display is output from the BCD
output.
The BCD output changes, too, when the display is changed with a
below mentioned keys.
: Change the value of changed digit.
: Select the changed digit.
: Finish the check mode.
When the BCD output is not installed, the BCD output is not confirmed.
There is BCD output
Move to the Confirmation of
ROM version
Confirmation of BCD
output
Confirmation of analog
output
There is analog
Confirmation of ROM
Not BCD output, Analog
なし
Summary of Contents for CSD-903-EX
Page 1: ...DIGITAL INDICATOR CSD 903 EX Instruction Manual EN294 1492 H ...
Page 2: ...IV ...
Page 91: ...75 ...
Page 93: ...77 ...
Page 95: ...79 ...
Page 97: ...81 ...
Page 199: ...183 Inform us about the contents of failure and situation at site in details ...
Page 204: ...188 Inform us about the contents of failure and situation at site in details ...
Page 206: ...190 Inform us about the contents of failure and situation at site in details ...
Page 209: ...193 Inform us about the contents of failure and situation at site in details ...
Page 224: ... ...
Page 225: ......