134
Error outside the setting range
When the data outside the setting range is written, the following code is set.
Condition Error
code
Contents
illegal data
0x03
When data outside [1] is written.
(4) Changing to calibration mode at zero point only
The calibration data such as zero point and the span point that is registered temporary are stored in
holding register address 56848 and returned to the measurement mode.
[1] is written in holding register address 56848.
[1] is automatically rewritten as [0] after writing is normally completed.
Error outside the setting range
When the data outside the setting range is written, the following code is set.
Condition Error
code
Contents
illegal data
0x03
When data outside [1] is written.
(5) Calibration forced finish.
When [0] is written in holding register address 51 during the calibration procedure, the calibration is
discontinued without storing the calibration data that is registered temporary, and you can return to
the measurement mode.