123
15-1. Calibration by transmission through Modbus interface
Please proceed the setting according to the following flow chart in this calibration.
15-1-1. Calibration mode
Each item of the flow chart corresponds to each procedure of [4-3. Calibration procedure details].
Please refer to [4-3. Calibration procedure details] for the details.
・
While executing the calibration by transmission,
(–RS–) is displayed.
・
When the calibration is compulsorily finished, all the data returns to the condition before the
calibration without executing the registration.
・
Please register the data for the calibration with the weight after confirming the stable condition
without fail.
(1) Change to Calibration mode by
communication
(2) Setting of Scale interval
(3) Setting of Weighing capacity value
(4) Setting of mass of weight
(7) ZERO calibration by numeric input
(10) SPAN calibration by numeric input
(5) Confirmation of stable at ZERO
(6) ZERO calibration by measuring value
(8) Confirmation of stable of SPAN
(9) SPAN calibration by weight
(11) Calibration is finished
In the calibration by numeric input
(Electrical calibration)
In the calibration by actual weight
(Calibration by actual weight)
Summary of Contents for CSD-903-EX
Page 1: ...DIGITAL INDICATOR CSD 903 EX Instruction Manual EN294 1492 H ...
Page 2: ...IV ...
Page 91: ...75 ...
Page 93: ...77 ...
Page 95: ...79 ...
Page 97: ...81 ...
Page 199: ...183 Inform us about the contents of failure and situation at site in details ...
Page 204: ...188 Inform us about the contents of failure and situation at site in details ...
Page 206: ...190 Inform us about the contents of failure and situation at site in details ...
Page 209: ...193 Inform us about the contents of failure and situation at site in details ...
Page 224: ... ...
Page 225: ......