Setting zero tracking time width
Initial value
=
00020
00000
=
Zero tracking OFF
Setting range
:
00000
〜
00099
Unit
:
0.1 s
Time width of 9.9 s with the setting “00099”.
※
Effective only when the value of 00001
〜
00099 is set with the F
−
08.
D
F
−
10
Setting of the target of HOLD
Initial value
=
01111
10
0
digit
:
Load display
10
1
digit
:
Comparator S1, S2 contact output, LED display
10
2
digit
:
Analog output
10
3
digit
:
Optional output
※
Out of the target at the setting “0”, and the target of HOLD is available at the setting
“1”.
D
F
−
11
Setting the CHECK value
Initial value
=
00004
Setting range
:
00000
〜
00014
Unit
:
0.2 mV/V(Setting can be made by every 0.2 mV/V)
Approx. 1.4 mV/V CHECK value at the setting “00014”.
D
F
−
12
Setting bridge power supply voltage
Initial value
=
00000
00000
:
5 V
00001
:
2.5 V
D
F
−
15
Setting digital filter for stabilized filter
Initial value
=
00003
00000
=
Stabilized filter OFF
Setting range
:
00000
〜
00005
Average of 2
m
times
※
Selects average times of digital filter for stabilized filter.
When the figure grows larger, the filter will be stronger, then the effects from
vibrations and so on are scarcely shown on the display.
※
Effective only when setting is made with the value from 00001 to 00999 with the
F
−
16, and the value from 00001 to 00999 is set with the F
−
17.
D
F
−
16
Setting time width for stabilized filter
Initial value
=
00020
00000
=
Stabilized filter OFF
Setting rage
:
00000
〜
00999
Unit
:
0.01 s
Time width of 9.99 s at the setting of “00999”.
※
Effective only when the value from 00001 to 00005 with the F
−
15 and the value from
00001 to 00999 with the F
−
97 are set.
D
F
−
17
Setting data width for stabilized filter
Initial value
=
00020
00000
=
Stabilized filter OFF
Setting rage
:
00000
〜
00999
Unit
:
1D
Data width of 999D at the setting of “00999”.
※
Effective only when the value from 00001 to 00005 with the F
−
15 and the value from
00001 to 00999 with the F
−
16 are set.