25
Procedures
1
Press the
key for about one second
The load display will show “FUNC”.
2
Press the
key twice.
It will make the load display proceed as
“FUNC”
→
“CCAL”
→
“ACAL”.
3
Press the
key.
“ACAL” mode can be entered, then the load
display will show “SCAL”.
4
Press the
key.
The load display shows “D
−
01”.
When the calibration has completed already, the
set value of minimum scale which has registered
at that time will be displayed.
Set the minimum scale with the right keys.
Setting value for the minimum scale will be 4
(four) as follows
:
1, 2, 5, 10
key
:
Set value inclement key
5
Press the
key.
The load display will show “DISP”.