23
Procedures
Press the
key.
The load display will show “ZERO” with lighting
display on and off, then zero adjustment can be
started.
10
Warning
:
At the same time, take care not
to apply load variation due to vibration and
so on. When load variation is applied, there
will be possibilities that zero point is
unstable, and precise reading of zero will
not be obtained.
After completed, the load display will become
“END”.
However, when the initial load is not entered
within the range from
−
0.3 mV/V to 2.4 mV/V,
the error code shown in the right figure will
show for about 2 seconds, then load display will
show “ZERO” and return to step 9.
Error Code
TE
−
L
:
Zero point
−
OVER
TE
−
H
:
Zero point
+
OVER
Error Code
Press the
key.
After “CCAL” mode is over, the load display will
h
th
t l
d
11
show the present load.