105
2
Reading
−
out the comparative data
(
Host
→
CSD
−
701B
)
Command No.
Operation
31
Reading out S1 data
32
Reading out S2 data
Return(CSD
−
701B
→
Host)
2
4
6
8
10
12
14
@
0
0
3
1
Terminator
Command No.
Sign
Setting value
ID No.
00
〜
31
D
The comparative data enter from the right end.
D
When the data is minus, “
−
”sign and when plus, “
+
” sign is added.
D
The sign data perform zero suppress.
D
When the decimal point is set with the F
−
56 and at the same time the decimal
point is determined with the F
−
01, it will be added to the specified position.
D
The empty sections are all spaces.
3
Writing comparative data
(
Host
→
CSD
−
701B
)
2
4
6
8
10
12
14
@
0
0
6
1
Terminator
Command No.
Sign
Set value
ID No.
00
〜
31
Command No.
Operation
61
Writing S1 data
62
Writing S2 data
Return(CSD
−
701B
→
Host)
2
4
6
8
10
12
14
1
6
@
0
0
6
1
Terminator
Command No.
Sign
Set value
ID No.
00
〜
31
D
The comparative data enters from the right end.
D
The setting range is within
−
99 999 to 99 999.
D
Don’t add a decimal point.
2
4
6
@
0
0
3
1
Terminator
Command No.
ID No.
00
〜
31