19
●
The accuracy of calibration obtained from
①
and
②
is 1/1 000 or so.
If more than the accuracy 1/1 000 is required, make calibration of
③
type.
In the following paragraphs, we will describe each calibration procedure by showing the
examples with load cell applied.
5
−
2
−
1. Calibration method to register the output of strain gage applied transducer at the time
of maximum display after setting the load to zero.
Warning
●
Before using a new instrument or exchanging the strain gage applied
transducer for a new one, be sure to make calibration.
If calibration shall not be made, correct measured results may not be
obtained nor it may cause malfunction in the instrument and there
may exist damage to the peripheral equipment.
Besides, even though the calibration has been made, there may occur
the similar case when the result is not correct, so make calibration
again.
●
During the calibration is executing, be sure to set Tare weight
cancellation clear, and to make cancellation (Execution of F
−
98) for
compensated data on zero set and set the OFF position of Zero
tracking(Setting “00000” on F
−
07 and F
−
08), and also set the OFF
position o Peak.
●
During calibration procedures, press the
key in case of
interrupting the calibration is required. The calibration data will be
kept as they are before entering the calibration and then returns to
the Measurement mode.
●
Every time the
key is pressed with the load display of ”FUNC”,
the display will change as the following arrow marks. However, every
time the
key is pressed, the display will change as the reverse
direction of the following arrow marks.
“FUNC”
→
“CCAL”
→
“ACAL”
→
“LCAL”
→
“ZERO”
→
“SPAN”
→
“TARE”
→
“TARE”
→
“CHEK”
→
“MONT”
→
“VCAL”
→
“VADJ”
→
“FUNC”
→
“CCAL”
→・・・・・
(Hereinafter, it will repeat.)