31
Procedures
Press the
key.
The load display shows “2000” and the minimum
display digit will flash on and off.
When the calibration has completed already, the
maximum display value which has registered at
that time will be displayed.
6
By the setting of minimum scale, the digit
of minimum display that flashes on and off
will be as follows
:
The minimum scale
1,2,5
10
0
digit
The minimum scale
10
10
1
digit
Set the maximum display value with the right
keys. Setting range for the maximum display
value will be (the minimum scale
×
100)
〜
99 990.
In order to make effective use of the performance,
set within the following ranges.
When setting is made over the range as below,
there may have a possibility of unstable display
and so on.
key
:
Set value carry key
key
:
Set value inclement key
key
:
Set value initialization key
Setting range for the
maximum display value
100
〜
10 000
200
〜
20 000
500
〜
50 000
1 000
〜
99 990
The minimum scale
1
2
5
10
It can be increased continuously by the
continuity press of the key.
7
Press the
key.
The load display will show “LOAD”.