Display
Cause
Solution
No display elements
No power present
Check power supply
-------
Key has no function in this status
Flashing
k
Battery defective or time changed
Set time
H
Load exceeds weighing capacity
Unload the scale
L
or
ERR 54
Weighing pan is not in place
Position the weighing pan
ERR 101 - 104
Key is stuck
Release key or
key pressed when switching on the device
Contact your local Service Center
ERR 320
Operating program memory faulty
Contact your local Service Center
ERR 335
Verified weighing platform not compatible
Connect a compatible weighing platform
with the connected terminal
ERR 340
Incorrect operating parameter (EEPROM)
Turn the scale off and then on again.
If the error code Err340 is still displayed
,
please contact your local Service Center
ERR 341
RAM has lost data
Connect the device to power for at least 10 hours
Battery is empty
ERR 343
Loss of data in the memory area for transaction Contact your local Service Center
numbers in external alibi memory
INF 01
“
Display overload
“
Data output not compatible with output format Set output format correctly
INF 02
“
Zero point error
Adjustment condition was not maintained,
Unload the scale first then zero, then tare via the
)
key
at start of cal.
“
e.g. not tared or weighing pan loaded
INF 03
Adjustment could not be completed
Allow to warm up again and repeat the adjustment
within a certain time.
process
INF 06
“
Int. weight not
Built-in adjustment weight is defective
Contact your local Service Center
available or missing
“
INF 07
“
Function not allowed
Function not allowed in scales verified for
Contact your local Service Center for
for verifiable weighing
“
use in legal metrology
information on changing settings
INF 08
The load on the scale is too heavy to zero
Check whether “Tare/zero at power on” was complied
the readout
with in your configuration.
INF 09
Taring is not possible when the scale gross
Zero the scale
weight is < zero
INF 10
Tare key is blocked when there is data in the
The data stored for the application program must be .
tare memory
deleted before taring
INF 22
Error in storing reference value
Load is too light, place a heavier weight on the weigher
INF 23
“
Application
Error in initializing an application
Contact your local Service Center
initialization error
“
INF 29
“
Scale minimum load
Minimum load not reached
Reduce minimum load (in the “Application parameters,”
not reached
“
under “Minimum load for autom. initialization”)
INF 71
“
Invalid value
,“
Cannot store the current weight value
None
Value too low
/
high
“ or
(e.g., control limits too low or too high)
“
selection is
not possible
“
INF 72
“
Maximum quantity
reached
“
Cannot store the current weight value
None
(e.g. transaction counter maximum reached)
INF 73
“
Memory deleted
/
Data not found or unreadable
Contact your local Service Center
Memory not
available
“
INF 74
“
Function is not
available
“ Function is blocked
None
or
(e.g. menu is blocked)
“
Function is blocked
“
INF 88
“
Function was
started
“
A function has been activated
None
INF 98
No weighing platform connected
Connect a weighing platform
INF 99
No weighing platform connected
Connect a weighing platform
NO WP
No weighing platform connected
Connect a weighing platform
116
Operating Instructions Combics Complete Scales
Error Codes
Summary of Contents for CAW3P
Page 125: ...Operating Instructions Combics Complete Scales 125 Specifications CAS G FE ...
Page 126: ...126 Operating Instructions Combics Complete Scales Specifications ...
Page 127: ...Operating Instructions Combics Complete Scales 127 Specifications CAS G IG ...
Page 128: ...128 Operating Instructions Combics Complete Scales Specifications ...
Page 134: ...34 Operating Instructions Combics Complete Scales Specifications ...
Page 138: ...138 Operating Instructions Combics Complete Scales Declarations of Conformity ...
Page 139: ...Operating Instructions Combics Complete Scales 139 Declarations of Conformity ...
Page 140: ...40 Operating Instructions Combics Complete Scales Declarations of Conformity 1 5 ...
Page 143: ...Operating Instructions Combics Complete Scales 143 EU type examination certificate ...
Page 144: ...144 Operating Instructions Combics Complete Scales EU type examination certificate ...
Page 145: ...Operating Instructions Combics Complete Scales 145 Test Certificate ...
Page 146: ...146 Operating Instructions Combics Complete Scales 146 Evaluation Certificate ...
Page 147: ...Operating Instructions Combics Complete Scales 147 14 Evaluation Certificate ...
Page 148: ...148 Operating Instructions Combics Complete Scales Evaluation Certificate ...
Page 149: ... Operating Instructions Combics Complete Scales 149 Certificate ...
Page 153: ...Operating Instructions Combics Complete Scales 153 ...
Page 154: ...154 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 155: ...Operating Instructions Combics Complete Scales 155 Plates and Markings ...
Page 156: ...156 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 168: ...168 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 169: ...Operating Instructions Combics Complete Scales 169 Plates and Markings ...
Page 170: ...170 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 171: ...Operating Instructions Combics Complete Scales 171 Plates and Markings ...
Page 172: ...172 Operating Instructions Combics Complete Scales ...
Page 173: ...Operating Instructions Combics Complete Scales 173 Certificate ...
Page 174: ...174 Operating Instructions Combics Complete Scales ...
Page 180: ...180 Operating Instructions Combics Complete Scales Certificate US C ...
Page 181: ...Operating Instructions Combics Complete Scales 181 Certificate US C ...
Page 182: ...182 Operating Instructions Combics Complete Scales ...
Page 184: ...184 Operating Instructions Combics Complete Scales ...
Page 185: ...Operating Instructions Combics Complete Scales 185 ...
Page 186: ...186 Operating Instructions Combics Complete Scales ...
Page 187: ...Operating Instructions Combics Complete Scales 187 ...