6-44 VS 9/VS 8 Vital Signs Monitor Service Manual
2. Flip up the black snap on the head of the display socket marked by 1 in the figure
and remove the display FPC cable; if a touchscreen is used, unplug the touchscreen
cable socket 2 to remove the main board.
6.4.7.5
(E-4-1) Removing the Encoder
Unscrew the 2 ST3.3*8 tapping screws 1, and slowly lift the board in the area with no
device on the board or directly clamp the knob with pliers and unplug the knob to
replace the encoder board and knob separately.