6-26 VS 9/VS 8 Vital Signs Monitor Service Manual
3. If a battery is loaded, unplug the battery socket and take out the battery (only one
socket 1 for the small-capacity battery and two socket 1 and 2 for large-capacity
batteries). Skip this step if no battery is loaded.
4. Remove the tape in the battery compartment. Remove the body temperature cable
from the socket as shown in the figure, and then the cable from the rear cover
buckle. (If the body temperature cable is not faulty, you do not need to remove the
cable. You only need to replace the body temperature module assembly to improve
maintenance efficiency).