6-28 VS 9/VS 8 Vital Signs Monitor Service Manual
6.4.2.2
When Recorder is Used.
1. Use a Phillips screwdriver to unscrew the 2 M3*6 pan head screws with washers at
the position shown in the figure on the rear cover.
2. Pinch the two lugs on the inside of the recorder as shown in the figure and press in
the direction of the arrow to remove the recorder.
3. Unplug the recorder cable to replace the recorder module or recorder cable only.